Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add filters








Language
Year range
1.
International Journal of Laboratory Medicine ; (12): 2789-2791,2793, 2015.
Article in Chinese | WPRIM | ID: wpr-602485

ABSTRACT

Objective To compare the performance of HP-083/4 and rational used instrument on detecting six items.Methods The rational instruments were used as contrast instrument,HP-083/4 was the verified instrument.A total of 100 blood specimens and 100 urine specimens were collected,and the levels of antistreptolysin O(ASO),hypersensitive C reactive protein (hsCRP),D-di-mer(D-D),glycosylated hemoglobin(HbA1c),rheumatoid factor(RF)and urine microalbumin(mAlb)were detected.The regression equation and correlation coefficient(r)of the two methods were calculated,and the Kappa values(κ)were analyzed to evaluate the performance of HP-083/4.Results There was a good linear correlation (r >0.950)for the two methods in detecting the serum ASO,hsCRP,D-D,HbA1c,RF and mAlb,r were 0.991,0.995,0.970,0.957,0.980 and 0.967 respectively.Besides,they had good concordance(κ>0.6),theκ values were 0.830,0.957,0.601,0.720,0.920 and 0.694 respectively.Conclusion HP-083/4 is effec-tive in detecting ASO,hsCRP,D-D,HbA1c,RF and mAlb,which should be suitable for clinical application.

SELECTION OF CITATIONS
SEARCH DETAIL